Artifact

Also called artefact, motion artifact, artifacts

An error in a heartbeat recording, such as a missed, extra or mistimed beat, often caused by movement or poor sensor contact. Artifacts can badly distort HRV numbers.

HRV is calculated from a list of beat-to-beat intervals, so it is only as good as that list. An artifact is any interval that does not reflect a real, normal heartbeat. Researchers split them into two kinds:

  • Technical artifacts come from the equipment. A loose electrode, a strap that shifts or a wrist that moves can make the sensor miss a beat, count one twice or time it wrongly.
  • Physiological artifacts come from the heart itself, mainly ectopic beats Ectopic beat A heartbeat that starts outside the heart's normal pacemaker and usually comes early. Ectopic beats are common, but they distort HRV unless removed or corrected. More about Ectopic beat that start outside the normal pacemaker Sinus node The heart's natural pacemaker, a small patch of cells that starts every normal heartbeat. Left alone it fires about 100 times a minute; the vagus nerve holds it lower at rest. More about Sinus node .

Either way, the result is an interval that is far too short or far too long. HRV metrics that compare neighboring beats, such as RMSSD RMSSD Root mean square of successive differences: an HRV metric built from how much each heartbeat interval differs from the next, in milliseconds. It mainly reflects vagal activity. More about RMSSD , are especially sensitive, because one bad interval creates two large differences. Lipponen and Tarvainen note that missed, extra and mistimed beat detections can significantly distort HRV values.

What it looks like. In a tachogram Tachogram A line chart of heartbeat intervals in the order they happened, one point per beat, in milliseconds. It shows HRV as a wavy line instead of a single number. More about Tachogram , the chart of beat-to-beat intervals, each kind of artifact leaves its own shape. A missed beat makes one interval about twice as long as its neighbours. A false detection splits one interval into two short ones. Movement scatters a run of values far above and below the normal line. These shapes are what cleaning rules look for.

Why one matters so much. Modelling missed and false detections, Berntson and Stowell (1998) found that a single artifact in about two minutes of data can add more false variability than the person’s real resting variability. In a short reading, one missed beat can multiply RMSSD several times over. When too many beats need fixing, a careful device or app refuses to report a number: the Kubios HRV guide suggests keeping corrected beats Beat correction Finding the beat intervals that are wrong (missed, extra or ectopic beats) and deleting or replacing them before HRV is calculated. The rule of thumb is to correct under 5% of beats. More about Beat correction under 5%, and this site’s chest-strap test stops at 10%. That is also why watches and rings take HRV while you are still or asleep, and why some nights show no HRV at all.

Good software finds and corrects artifacts before calculating HRV, either by deleting the bad intervals or by replacing them with estimated values. Different methods can give slightly different results, so the cleaning step is one of several reasons devices disagree.

Plant a missed, extra or ectopic beat in the five-minute teaching recording below, then try each correction method and compare RMSSD, SDNN SDNN The standard deviation of all NN intervals in a recording, in milliseconds. It measures overall variability, and its value depends strongly on how long the recording lasts. More about SDNN and the spectrum with the clean recording.

Clean one bad beat

The teaching recording, 341 intervals

Flagged: 1 of 341 intervals

Measure Clean Bad beat Fixed
RMSSD (ms) 33.4 70.2 70.2
SDNN (ms) 28.7 51.7 51.7
pNN50 (%) 8.2 8.8 8.8
LF power (ms²) 120 1241 1241
HF power (ms²) 635 2314 2314

A missed beat makes one interval about twice as long as its neighbours. Left in, it lifts RMSSD from 33.4 to 70.2 ms, 2.1 times the true value, while SDNN moves less.

The buttons work when JavaScript is on.

One bad beat in the five-minute teaching recording. The rule flags any interval more than 250 ms from the median of its neighbours; the method you pick repairs it. Compare the numbers before and after with the clean recording.

A line of the gap between beats over the first 60 seconds of the teaching recording, with the planted artifact as a spike, a circle on each flagged interval and the corrected line drawn over it. A table gives RMSSD, SDNN, pNN50, LF power and HF power for the clean recording, with the artifact left in, and after correction, plus how many intervals were flagged. One group of buttons plants a missed beat, an extra beat or an ectopic beat, or restores the clean recording; another chooses the correction method: leave it in, delete, mean of neighbours, linear interpolation or cubic spline. A missed beat left in roughly doubles RMSSD; every method brings it back within a few percent.

You can prevent many artifacts yourself. Keep still during a spot reading, wear the strap or band snugly, and wet chest-strap electrodes before use. If one reading is far from your usual range, it is worth retaking before you read meaning into it.

The lesson How to take a good HRV reading gives a step-by-step routine, and Artifacts and extra beats shows every pattern, its everyday causes and the correction methods side by side.

Sources

  1. Role of editing of R-R intervals in the analysis of heart rate variability Frontiers in Physiology, 2012
  2. A robust algorithm for heart rate variability time series artefact correction using novel beat classification Journal of Medical Engineering & Technology, 2019
  3. ECG artifacts and heart period variability: Don't miss a beat! Psychophysiology, 1998
  4. Kubios HRV User's Guide, version 3.5 Kubios Oy, 2021

All glossary terms

All HRV parameters